<p>The current study aimed to investigate the feasibility of using the pseudo-blood<sup>®</sup>, adjusted to equal specific gravity of beads and solution, as a phantom for diffusion-weighted imaging (DWI) evaluation. The pseudo-blood<sup>®</sup> with bead concentrations was adjusted to 0%, 9.8%, 28.1%, and 48.5% and was used as phantoms. Approximately 3-min of DWI was performed 15 times at 10-min intervals. The same experiment was conducted after 30, 60, and 90 days. The coefficients of variation of the obtained intra- and inter-day apparent diffusion coefficient (ADC) values were calculated. The ADC values significantly decreased with increasing bead concentrations in the pseudo-blood<sup>®</sup>, with ADC values of 0.789 ± 0.002, 0.753 ± 0.001, 0.709 ± 0.003, and 0.685 ± 0.003 × 10<sup>−3</sup>&#xa0;mm<sup>2</sup>&#xa0;s<sup>−1</sup> at bead concentrations of 0%, 9.8%, 28.1%, and 48.5% (Day 0), respectively. The intra- and inter-day coefficients of variation and repeatability coefficient were &lt; 5% and &lt; 5 × 10<sup>−5</sup>&#xa0;mm<sup>2</sup>&#xa0;s<sup>−1</sup> for all phantoms. Therefore, the pseudo-blood<sup>®</sup> can be used as a phantom for DWI evaluation.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Evaluation of the Short- and Long-Term Stability of Apparent Diffusion Coefficient Using the Commercial Product Pseudo-Blood®

  • Kazuya Sakoda,
  • Shogo Baba

摘要

The current study aimed to investigate the feasibility of using the pseudo-blood®, adjusted to equal specific gravity of beads and solution, as a phantom for diffusion-weighted imaging (DWI) evaluation. The pseudo-blood® with bead concentrations was adjusted to 0%, 9.8%, 28.1%, and 48.5% and was used as phantoms. Approximately 3-min of DWI was performed 15 times at 10-min intervals. The same experiment was conducted after 30, 60, and 90 days. The coefficients of variation of the obtained intra- and inter-day apparent diffusion coefficient (ADC) values were calculated. The ADC values significantly decreased with increasing bead concentrations in the pseudo-blood®, with ADC values of 0.789 ± 0.002, 0.753 ± 0.001, 0.709 ± 0.003, and 0.685 ± 0.003 × 10−3 mm2 s−1 at bead concentrations of 0%, 9.8%, 28.1%, and 48.5% (Day 0), respectively. The intra- and inter-day coefficients of variation and repeatability coefficient were < 5% and < 5 × 10−5 mm2 s−1 for all phantoms. Therefore, the pseudo-blood® can be used as a phantom for DWI evaluation.