Development of Magnetic Resonance Imaging (MRI) Phantom With Arbitrary T1 Value Using Dextrin Hydrate
摘要
To formulate the relationship between the dextrin hydrate concentration and T1 relaxation rate and create a magnetic resonance imaging (MRI) phantom with an arbitrary T1 value. Dextrin solution with nine different concentrations was prepared by dissolving 0–20 g (2.5 g increments) of dextrin in 25 g of purified water. The T1 values of the phantoms were measured using a 1.5 T MR scanner, and the relationship between the R1 value and dextrin concentration was regressed using linear and quadratic equations. Phantoms with concentrations adjusted to T1 values of 500, 1000, and 1500 ms were created from each regression equation, and the errors between the measured T1 and target values were evaluated. In addition, the temporal changes in the T1 and T2 values of the phantoms were also evaluated. The T1 and T2 values ranged from 367.4 ± 14.1 to 2577.6 ± 76.5 ms and 20.0 ± 0.9 to 1805.3 ± 8.3 ms, respectively. The linear and quadratic regression equations were