Fluorescence turn-off detection of myoglobin as a cardiac biomarker using water-stable L-glutamic acid functionalized cesium lead bromide perovskite quantum dots

  • Bhargava Thamminana,
  • Mayurkumar Revabhai Patel,
  • Madhura Pradeep Deshpande,
  • Tae Jung Park,
  • Suresh Kumar Kailasa

摘要

Water dispersible L-glutamic acid (Glu) functionalized cesium lead bromide perovskite quantum dots (CsPbBr3 PQDs), namely CsPbBr3@Glu PQDs were synthesized and used for the fluorescence “turn-off” detection of myoglobin (Myo). The as-prepared CsPbBr3@Glu PQDs exhibited an exceptional photoluminescence quantum yield of 25% and displayed emission peak at 520 nm when excited at 380 nm. Interestingly, the fluorescence “turn-off” analytical approach was designed to detect Myo using CsPbBr3@Glu PQDs as a simple optical probe. The developed probe exhibited a wide linear range (0.1–25 µM) and a detection limit of 42.42 nM for Myo sensing. The CsPbBr3@Glu PQDs-based optical probe provides high ability to determine Myo in serum and plasma samples.