Purpose <p>To evaluate the feasibility of secure telesurgery by assessing the impact of image compression and cybersecurity devices on surgical performance and data transmission.</p> Methods <p>Telesurgical procedures using the&#xa0;<i>hinotori™</i>&#xa0;surgical robot were performed remotely between Hirosaki and Goshogawara, which are 30&#xa0;km apart, over a secure line provided by NTT East. Image compression was tested at 120, 80, 60, 40, and 20&#xa0;Mbps. A surgical specialist operated on artificial organ models. Simulated cyberattacks were introduced to assess the performance of security devices.</p> Results <p>Even at 20&#xa0;Mbps, there was no significant loss in operability or image quality. Security devices detected simulated attacks and permitted essential robot communications. No visual distortion or operational issues occurred, and only a small delay of ≤ 2&#xa0;min was introduced. The transmission control protocol (TCP) error rates remained low, with or without security devices (0.00–0.04%).</p> Conclusion <p>Security implementation enables safe telesurgery by detecting cyber threats without impairing surgical performance. This finding supports the practical development of economically and technologically viable telesurgical systems.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Impact of data compression and security devices on telesurgery systems

  • Hajime Morohashi,
  • Kenichi Hakamada,
  • Yusuke Wakasa,
  • Kazuki Yokoyama,
  • Yuma Ebihara,
  • Satoshi Hirano,
  • Eiji Oki,
  • Norihiko Ikeda,
  • Akinobu Taketomi,
  • Masaki Mori

摘要

Purpose

To evaluate the feasibility of secure telesurgery by assessing the impact of image compression and cybersecurity devices on surgical performance and data transmission.

Methods

Telesurgical procedures using the hinotori™ surgical robot were performed remotely between Hirosaki and Goshogawara, which are 30 km apart, over a secure line provided by NTT East. Image compression was tested at 120, 80, 60, 40, and 20 Mbps. A surgical specialist operated on artificial organ models. Simulated cyberattacks were introduced to assess the performance of security devices.

Results

Even at 20 Mbps, there was no significant loss in operability or image quality. Security devices detected simulated attacks and permitted essential robot communications. No visual distortion or operational issues occurred, and only a small delay of ≤ 2 min was introduced. The transmission control protocol (TCP) error rates remained low, with or without security devices (0.00–0.04%).

Conclusion

Security implementation enables safe telesurgery by detecting cyber threats without impairing surgical performance. This finding supports the practical development of economically and technologically viable telesurgical systems.