Purpose <p>This study aimed to analyze adverse events associated with Mazor X surgical guidance robots, focusing on parameters such as root causes, vertebral distribution, surgical delay, robot abandonment, and more.</p> Methods <p>330 adverse events reported within the Manufacturer and User Facility Device Experience (MAUDE) database for the Mazor X (MX) and Mazor X Stealth Edition (MXSE) robot guidance systems were analyzed and categorized based on the root causes, vertebral distribution, surgical delay, robot abandonment, and patient outcomes.</p> Results <p>Both the MX and MXSE adverse events were primarily due to external root causes (35% and 52%, respectively), with user error being the most common. Deviations were most likely to affect the L3, L4, and L5 vertebra, and deviate either medially or laterally. Motor deficits were the most common adverse patient outcome for MX, while sensory deficits were the most common for MXSE. There were two deaths associated with the MX, compared to zero deaths with the MXSE.</p> Conclusion <p>While the MAUDE database cannot be used to calculate the incidence or prevalence of adverse events, it provides insight into the types of adverse events that tend to occur with medical devices. MAUDE reports suggest that user-related errors account for most of the adverse events that occur, underscoring the need for focused surgeon training. Our study highlights the importance of understanding both the types of adverse events and their root causes to promote patient safety and reduce complication rates.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Analysis of root causes of adverse events with Mazor X and Mazor X stealth edition-guided spine surgeries: an FDA MAUDE database study

  • Apratim Maity,
  • Ethan D. L. Brown,
  • Max Ward,
  • Emily A. Orsino,
  • Daniel M. Schneider,
  • Daniel M. Sciubba,
  • Sheng-fu Larry Lo

摘要

Purpose

This study aimed to analyze adverse events associated with Mazor X surgical guidance robots, focusing on parameters such as root causes, vertebral distribution, surgical delay, robot abandonment, and more.

Methods

330 adverse events reported within the Manufacturer and User Facility Device Experience (MAUDE) database for the Mazor X (MX) and Mazor X Stealth Edition (MXSE) robot guidance systems were analyzed and categorized based on the root causes, vertebral distribution, surgical delay, robot abandonment, and patient outcomes.

Results

Both the MX and MXSE adverse events were primarily due to external root causes (35% and 52%, respectively), with user error being the most common. Deviations were most likely to affect the L3, L4, and L5 vertebra, and deviate either medially or laterally. Motor deficits were the most common adverse patient outcome for MX, while sensory deficits were the most common for MXSE. There were two deaths associated with the MX, compared to zero deaths with the MXSE.

Conclusion

While the MAUDE database cannot be used to calculate the incidence or prevalence of adverse events, it provides insight into the types of adverse events that tend to occur with medical devices. MAUDE reports suggest that user-related errors account for most of the adverse events that occur, underscoring the need for focused surgeon training. Our study highlights the importance of understanding both the types of adverse events and their root causes to promote patient safety and reduce complication rates.