Our estimates of neonatal radiation exposure fall short of reality
摘要
The purpose of this study is to evaluate radiation exposure in newborns undergoing imaging tests during the first 30 days of neonatal intensive care unit (NICU) hospitalization. A retrospective cohort study was conducted from November 2018 to April 2019 with newborns admitted to the NICU. Thermoluminescent dosimeters (TLD-100™) measured radiation emitted during imaging exams over 1 month, with a comparison between measured and estimated radiation. The cohort exhibited a median gestational age of 33.0 (31.0, 37.0) weeks, a median birth weight of 1840 (1272, 2748) g, and a median length of stay of 25.5 (11.7, 55.0) days. Eighty-four patients underwent 314 imaging tests, with an estimated radiation dose (ERD) per patient of 0.116 mSv and a measured radiation dose (MDR) of 0.158 mSv. ERD consistently underestimated MDR, with a mean difference of −0.043 mSv (−0.049 to −0.036) in the Bland-Altman analysis. The regression equation was as follows: difference MRD – ERD = −1.7 × (mean (MRD + ERD)) + 0.056. The mean estimated radiation dose per exam was 0.030 mSv, and the chest X-rays accounted for 63.26% of total exams. The median number of radiographic incidences per patient was 2 (1, 4), with 5 patients undergoing three or more exams in a single day.
Conclusion: Radiation exposure in these newborns was underestimated, emphasizing the need for awareness regarding associated risks and strict criteria for requesting radiological exams. Lung ultrasound is a radiation-free and effective option in managing respiratory diseases in newborns, reducing the reliance on chest X-rays.