Purpose <p>Skin temperature is a critical component of athletes’ heat strain during exercise in the heat. This study evaluated the innovative Flex sensor (BodyCap, Caen, France), allowing real-time measurements, against a data logger (iButton hygrochron DS1923, Analog Devices, Wilmington, USA), commonly used in sports science.</p> Methods <p>In a first study, the accuracies of iButton faces as well as the Flex sensor were tested on a constant-temperature Peltier module affixed to a brass plate. In a second study, concurrent skin temperature measurements were conducted with both sensors on eight participants cycling in hot conditions.</p> Results <p>Both faces of the iButton underestimated plate temperature, with the base (ME = mean error: 0.59 ± 0.25&#xa0;°C at 30&#xa0;°C, 1.62 ± 0.48&#xa0;°C at 40&#xa0;°C) being significantly more accurate than the top (ME: 1.14 ± 0.26&#xa0;°C at 30&#xa0;°C, 3.02 ± 0.71&#xa0;°C at 40&#xa0;°C, <i>p</i> &lt; 0.001). The rate of temperature increase of the base was also higher than the top at 30&#xa0;°C (1.97 ± 0.16&#xa0;°C/min vs 1.49 ± 0.15&#xa0;°C/min) and 40&#xa0;°C (4.41 ± 0.46&#xa0;°C/min vs 3.14 ± 0.16 °C/min, both <i>p</i> &lt; 0.001). The Flex sensor underestimated plate temperature (ME: 1.31 ± 0.32&#xa0;°C at 30&#xa0;°C, 3.00 ± 0.84&#xa0;°C at 40&#xa0;°C), showing larger errors than the iButton base (<i>p</i> &lt; 0.001). During exercise, the Flex sensor (36.91 ± 1.19&#xa0;°C) measured higher skin temperature compared to the iButton (36.50 ± 1.15&#xa0;°C), with an ME of -0.41 ± 0.55&#xa0;°C.</p> Conclusion <p>In the absence of a gold standard, for scenarios where real-time monitoring is not required, using the iButton with its base applied to the skin is therefore recommended. To enhance the Flex sensor’s accuracy, implementing a calibration protocol before use is strongly suggested, enabling its full potential for real-time monitoring.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Validation of the iButton and Flex sensors for measuring skin temperature

  • David Bandiera,
  • Jean de Bardonnèche,
  • Philippe Janry,
  • Sylvie Calas-Étienne,
  • Jean-Christophe Aubin,
  • Antonio Tessitore,
  • Yannis Pitsiladis,
  • Sébastien Racinais

摘要

Purpose

Skin temperature is a critical component of athletes’ heat strain during exercise in the heat. This study evaluated the innovative Flex sensor (BodyCap, Caen, France), allowing real-time measurements, against a data logger (iButton hygrochron DS1923, Analog Devices, Wilmington, USA), commonly used in sports science.

Methods

In a first study, the accuracies of iButton faces as well as the Flex sensor were tested on a constant-temperature Peltier module affixed to a brass plate. In a second study, concurrent skin temperature measurements were conducted with both sensors on eight participants cycling in hot conditions.

Results

Both faces of the iButton underestimated plate temperature, with the base (ME = mean error: 0.59 ± 0.25 °C at 30 °C, 1.62 ± 0.48 °C at 40 °C) being significantly more accurate than the top (ME: 1.14 ± 0.26 °C at 30 °C, 3.02 ± 0.71 °C at 40 °C, p < 0.001). The rate of temperature increase of the base was also higher than the top at 30 °C (1.97 ± 0.16 °C/min vs 1.49 ± 0.15 °C/min) and 40 °C (4.41 ± 0.46 °C/min vs 3.14 ± 0.16 °C/min, both p < 0.001). The Flex sensor underestimated plate temperature (ME: 1.31 ± 0.32 °C at 30 °C, 3.00 ± 0.84 °C at 40 °C), showing larger errors than the iButton base (p < 0.001). During exercise, the Flex sensor (36.91 ± 1.19 °C) measured higher skin temperature compared to the iButton (36.50 ± 1.15 °C), with an ME of -0.41 ± 0.55 °C.

Conclusion

In the absence of a gold standard, for scenarios where real-time monitoring is not required, using the iButton with its base applied to the skin is therefore recommended. To enhance the Flex sensor’s accuracy, implementing a calibration protocol before use is strongly suggested, enabling its full potential for real-time monitoring.