Reliability and validity of the MX3 portable sweat sodium analyser during exercise in warm conditions

  • Harry A. Brown,
  • Brad Clark,
  • Julien D. Périard

摘要

Purpose

Accurately measuring sweat sodium concentration ([Na+]) in the field is advantageous for coaches, scientists, and dieticians looking to tailor hydration strategies. The MX3 hydration testing system is a new portable analyser that uses pre-calibrated biosensors to measure sweat [Na+]. This study aimed to assess the validity and reliability of the MX3 hydration testing system.

Methods

Thirty-one (11 females) recreationally active participants completed one experimental trial. During this trial, participants exercised at a self-selected pace for 45 min in a warm environment (31.5 ± 0.8 °C, 63.2 ± 1.3% relative humidity). Sweat samples were collected from three measurement sites using absorbent patches. The samples were then analysed for sweat [Na+] using both the MX3 hydration testing system and the Horiba LAQUAtwin-NA-11. The reliability of the MX3 hydration testing system was determined following two measurements of the same sweat sample.

Results

The mean difference between measurements was 0.1 mmoL·L−1 (95% limits of agreement (LoA): − 9.2, 9.4). The analyser demonstrated a coefficient of variation (CV) of 5.6% and the standard error of measurement was 3.3 mmoL·L−1. When compared to the Horiba LAQUAtwin-NA-11, there was a mean difference of − 1.7 mmoL·L−1 (95% LoA: − 0.25 \({\overline{\text{X}}}\) X ¯ , 0.25 \({\overline{\text{X}}}\) X ¯ ) and the CV was 9.8%.

Conclusion

The MX3 hydration testing system demonstrated very good single-trial reliability, moderate agreement and a very good CV relative to the Horiba LAQUAtwin-Na-11. To further validate its performance, the MX3 hydration testing system should be compared with analytical techniques known for superior reliability and validity.