Purpose <p>To compare corneal biomechanical properties in patients affected by Myotonic Dystrophy (DM1) with healthy subjects in order to better understand the underlying reasons for low intraocular pressure (IOP) values detected in DM1 eyes.</p> Methods <p>in this retrospective, comparative study were included 41 eyes of 41 DM1 patients and 41 eyes of healthy subjects, age and sex matched, were included. All participants underwent a complete eye visit including Goldmann applanation tonometry (GAT), corneal biomechanical properties evaluation performed with Corvis ST (CST) and corneal morphological evaluation through Oculus Pentacam.</p> Results <p>In this study, GAT-measured IOP was significantly lower in DM1 patients (12.63 ± 2.12 mmHg) compared to controls (16.65 ± 2.24 mmHg) (<i>p</i> &lt; 0.001). DM1 patients, compared to the healthy subjects, showed significant differences in many parameters related to the corneal deformation. In particular, DM1 patients displayed, among others, higher mean values of highest concavity deformation amplitude (1.19&#xa0;mm vs. 1.04&#xa0;mm) (<i>p</i> &lt; 0.0008), of deflection amplitude measured at applanation 2 (0.15&#xa0;mm vs. 0.11&#xa0;mm) (<i>p</i> &lt; 0.0008), deflection amplitude measured at highest concavity (0.98&#xa0;mm vs. 0.87&#xa0;mm) (<i>p</i> &lt; 0.0008) and maximum length of deflection amplitude (1.05&#xa0;mm vs. 0.88&#xa0;mm) (<i>p</i> &lt; 0.0008).</p> Conclusion <p>The findings presented in this study suggest that the corneas of patients affected by DM1 are more susceptible to external stimuli than those of healthy subjects. Thus, GAT may underestimate IOP in DM1 patients due to altered corneal biomechanics.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Intraocular pressure and corneal biomechanics in patients affected by myotonic dystrophy type 1

  • Michele Lanza,
  • Rosa Boccia,
  • Marianna Scutifero,
  • Luigi Serra,
  • Teresa Cangiano,
  • Esther Picillo,
  • Italo Fattore,
  • Vincenzo Nigro,
  • Francesca Simonelli

摘要

Purpose

To compare corneal biomechanical properties in patients affected by Myotonic Dystrophy (DM1) with healthy subjects in order to better understand the underlying reasons for low intraocular pressure (IOP) values detected in DM1 eyes.

Methods

in this retrospective, comparative study were included 41 eyes of 41 DM1 patients and 41 eyes of healthy subjects, age and sex matched, were included. All participants underwent a complete eye visit including Goldmann applanation tonometry (GAT), corneal biomechanical properties evaluation performed with Corvis ST (CST) and corneal morphological evaluation through Oculus Pentacam.

Results

In this study, GAT-measured IOP was significantly lower in DM1 patients (12.63 ± 2.12 mmHg) compared to controls (16.65 ± 2.24 mmHg) (p < 0.001). DM1 patients, compared to the healthy subjects, showed significant differences in many parameters related to the corneal deformation. In particular, DM1 patients displayed, among others, higher mean values of highest concavity deformation amplitude (1.19 mm vs. 1.04 mm) (p < 0.0008), of deflection amplitude measured at applanation 2 (0.15 mm vs. 0.11 mm) (p < 0.0008), deflection amplitude measured at highest concavity (0.98 mm vs. 0.87 mm) (p < 0.0008) and maximum length of deflection amplitude (1.05 mm vs. 0.88 mm) (p < 0.0008).

Conclusion

The findings presented in this study suggest that the corneas of patients affected by DM1 are more susceptible to external stimuli than those of healthy subjects. Thus, GAT may underestimate IOP in DM1 patients due to altered corneal biomechanics.