Keratometry repeatability in healthy and post-refractive surgery eyes: a comparison of two swept-source devices
摘要
To analyze the keratometry repeatability obtained with two swept-source devices in healthy and post-refractive surgery eyes.
MethodsOne hundred volunteers aged between 20 and 55 years were recruited for this study including both healthy and post-refractive surgery eyes. Patients included in the post-refractive surgery group underwent Laser-Assisted in Situ Keratomileusis (LASIK) or Photorefractive Keratectomy (PRK). Three consecutive measurements were taken with the IOLMaster 700 and Anterion and the evaluated measurements were simulated keratometry (Sim K), posterior keratometry (PK), and total keratometry (TK). Repeatability was assessed through the intra-subject standard deviation (Sw), repeatability limit (R), and variability coefficient (CoV).
ResultsFifty patients were included in the control and post-refractive surgery groups, respectively. Similar repeatability was obtained for healthy and post-refractive surgery eyes using both devices. For IOLMaster 700, R ranged between 0.16 and 0.22 D for Sim K, 0.09–0.11 D for PK, and 0.18–0.24 D for TK. For Anterion, R ranged between 0.19 and 0.20 D for Sim K, 0.02 D for PK, and 0.21-0.22 D for TK. In general, the control group showed slightly higher Sw and R values than the post-refractive surgery group for TK and astigmatism measurements obtained with both devices.