错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

MOG-IgG testing strategies in accordance with the 2023 MOGAD criteria: a clinical-laboratory assessment

  • Mario Risi,
  • Giacomo Greco,
  • Stefano Masciocchi,
  • Eleonora Rigoni,
  • Elena Colombo,
  • Pietro Businaro,
  • Silvia Scaranzin,
  • Chiara Morandi,
  • Alvino Bisecco,
  • Paola Bini,
  • Luca Diamanti,
  • Antonio Gallo,
  • Diego Franciotta,
  • Matteo Gastaldi

摘要

Background

Live cell-based assay (LCBA) is the gold standard for MOG-IgG detection, and fixed CBA (FCBA) is a widely used commercial alternative. Recent criteria attributed a diagnostic value to MOG-IgG titration with both LCBA and FCBA, with low-titre samples requiring additional supporting features for MOGAD diagnosis. However, FCBA titration is not validated. We aimed to assess the impact of the criteria-based MOG-IgG testing in MOGAD diagnosis.

Methods

Thirty-eight serum samples of LCBA MOG-IgG1-positive MOGAD patients were titred on MOG-IgG LCBA and FCBA, and the presence of supporting features for MOGAD assessed. MOGAD criteria were evaluated in four testing scenarios: (a) FCBA without titration; (b) FCBA with titration; c) LCBA without titration; (d) LCBA with titration.

Results

FCBA without titration failed to reach MOGAD diagnosis in 11/38 patients (28.9%, negative results in 5, lack of supporting features in 6). Patients with unconfirmed diagnosis had optic neuritis (ON, n = 8), or transverse myelitis (TM, n = 3). FCBA with titration allowed MOGAD diagnosis in 4 additional patients. Correlation between LCBA and FCBA titres was moderate (Spearman’s rho 0.6, p < 0.001).

Conclusions

FCBA yields high rate of misdiagnosis mainly due a lower analytical sensitivity. FCBA titration provides a moderate diagnostic advantage in FCBA positive patients.