Towards a phantom-based quantitative assessment of dose reduction and image-quality trade-offs in half- and full-rotation dental CBCT protocols: a thermoluminescent dosimetry approach with clinical insight

  • Hakan Amasya,
  • Osman Günay,
  • Fahrettin Fatih Kesmezacar,
  • Emine Tuna Akdoğan,
  • Duygu Tunçman Kayaokay,
  • Songül Çavdar Karaçam,
  • Nami Yeyin,
  • Mustafa Demir,
  • Kaan Orhan,
  • Ghada ALMisned,
  • H. O. Tekin

摘要

This study aimed to compare full- (360°) and half-rotation (180°) dental cone-beam computed tomography (CBCT) protocols in terms of effective dose (ED) and quantitative image quality for the maxilla and mandible regions. An Alderson Radiation Therapy phantom was imaged using a Hyperion X9 Pro (Cefla, Imola) CBCT device with a 10 × 6 cm field of view. Both anatomical regions were acquired using full- and half-rotation protocols. A total of 67 thermoluminescence dosimeters were positioned in the phantom for ED calculations. Quantitative image-quality assessment was performed using axial slices extracted from each volume, and signal-to-noise ratio (SNRs) and contrast-to-noise ratio (CNRs) were calculated. The EDs were 406.33 µSv (full-rotation) and 208.29 µSv (half-rotation) for the maxilla, and 248.94 µSv (full-rotation) and 73.63 µSv (half-rotation) for the mandible. These doses corresponded to dose reductions of 48.74% in the maxilla and 70.42% in the mandible. SNRs and CNRs decreased by 36.69% and 34.74% in the maxilla, and by 49.95% and 50.28% in the mandible, respectively. Dose reduction was more pronounced in the mandible than in the maxilla, and this was accompanied by a greater loss in SNR and CNR. It is concluded that protocol selection should be guided by diagnostic requirements and adherence to the “As Low as Diagnostically Acceptable being Indication-oriented and Patient-specific (ALADAIP) principle.