Study goal <p>The electrically evoked stapedius reflex threshold (eSRT) is widely used as an estimate of the maximum comfort level (MCL) in the fitting of cochlear implants (CI). This study investigated the long-term evolution of the eSRT during a ten-year observation period.</p> Methods <p>Retrospective analysis of fitting map data (charge units [QU] required to evoke the stapedius reflex) obtained from 50 cochlear-implanted ears of 26 children. Time-series statistical methods were employed to examine trends in charge unit (QU) values and their variance, normalized to individual baselines.</p> Results <p>The QU time series mean increased significantly within the first six months post-implantation and reached a plateau thereafter. Variance in eSRT decreased significantly over the first five years, with three key turning points identified at six months, 18&#xa0;months, and 4.25&#xa0;years, indicating significantly diminishing variation with time.</p> Discussion <p>CI fitting schedules should be coordinated with temporal changes in eSRT variance. Frequent fittings in the first 18&#xa0;months, followed by annual fittings until four years after implantation, seem advisable. After this period, less frequent fittings may be sufficient.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Longitudinal changes in electrical stapedius reflex thresholds (eSRT) in children with cochlear implants

  • Franz Muigg,
  • Simone Graf,
  • Maria Magdalena Mair,
  • Josef Seebacher,
  • Hazem Salloum,
  • Viktor Weichbold,
  • Joachim Schmutzhard,
  • Philipp Zelger

摘要

Study goal

The electrically evoked stapedius reflex threshold (eSRT) is widely used as an estimate of the maximum comfort level (MCL) in the fitting of cochlear implants (CI). This study investigated the long-term evolution of the eSRT during a ten-year observation period.

Methods

Retrospective analysis of fitting map data (charge units [QU] required to evoke the stapedius reflex) obtained from 50 cochlear-implanted ears of 26 children. Time-series statistical methods were employed to examine trends in charge unit (QU) values and their variance, normalized to individual baselines.

Results

The QU time series mean increased significantly within the first six months post-implantation and reached a plateau thereafter. Variance in eSRT decreased significantly over the first five years, with three key turning points identified at six months, 18 months, and 4.25 years, indicating significantly diminishing variation with time.

Discussion

CI fitting schedules should be coordinated with temporal changes in eSRT variance. Frequent fittings in the first 18 months, followed by annual fittings until four years after implantation, seem advisable. After this period, less frequent fittings may be sufficient.