Purpose <p>To investigate hearing preservation in patients who received a cochlear implant with a preformed electrode carrier, namely the Slim Modiolar (SM).</p> Methods <p>A monocentric, retrospective study was conducted. A total of 51 adult patients (24 females, 27 males), representing 60 ears, were included in the final evaluation.</p> <p>The mean age was 60.9 ± 15.2 years (range 17.5–89.7 years). All patients had some degree of residual hearing. Frequency-specific pure-tone audiometry was performed at three time points: preoperatively (T0), within 30 days postoperatively (T1), and at a later time point during follow-up (T2; 20.7 ± 17.6 months). Low frequency pure tone average and the HEARRING group formula were used to facilitate comparison of hearing preservation rates with other results published so far.</p> Results <p>In the first month after surgery, no significant deterioration was observed for the low frequencies, but a significant deterioration was observed at follow-up. Using the HEARRING classification system, the average short-term (T0–T1) hearing preservation score was 70%, which corresponds to partial hearing preservation. During follow-up (T0–T2), the score decreased to 61% in the study group, still representing partial hearing preservation. Complete hearing loss occurred in 20% of the cases.</p> Conclusion <p>Especially in the low frequency range, the utilisation of SM electrode carriers during cochlear implantation facilitates preservation of residual hearing.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Cochlear implantation with Slim Modiolar Electrode carriers enables hearing preservation

  • Kenan Dennis Sakmen,
  • Christian Issing,
  • Caterina Vazzana,
  • Tobias Weißgerber,
  • Albrecht Linke,
  • Timo Stöver,
  • Silke Helbig

摘要

Purpose

To investigate hearing preservation in patients who received a cochlear implant with a preformed electrode carrier, namely the Slim Modiolar (SM).

Methods

A monocentric, retrospective study was conducted. A total of 51 adult patients (24 females, 27 males), representing 60 ears, were included in the final evaluation.

The mean age was 60.9 ± 15.2 years (range 17.5–89.7 years). All patients had some degree of residual hearing. Frequency-specific pure-tone audiometry was performed at three time points: preoperatively (T0), within 30 days postoperatively (T1), and at a later time point during follow-up (T2; 20.7 ± 17.6 months). Low frequency pure tone average and the HEARRING group formula were used to facilitate comparison of hearing preservation rates with other results published so far.

Results

In the first month after surgery, no significant deterioration was observed for the low frequencies, but a significant deterioration was observed at follow-up. Using the HEARRING classification system, the average short-term (T0–T1) hearing preservation score was 70%, which corresponds to partial hearing preservation. During follow-up (T0–T2), the score decreased to 61% in the study group, still representing partial hearing preservation. Complete hearing loss occurred in 20% of the cases.

Conclusion

Especially in the low frequency range, the utilisation of SM electrode carriers during cochlear implantation facilitates preservation of residual hearing.