Bacterial biofilm formation on headpieces of Cochlear implants

  • Lisa Kirchhoff,
  • Diana Arweiler-Harbeck,
  • Moritz Meyer,
  • Jan Buer,
  • Stephan Lang,
  • Joerg Steinmann,
  • Ralf Bertram,
  • Eric Deuss,
  • Benedikt Höing

摘要

Introduction

Bacterial biofilm formation on medical devices, such as Cochlear implants (CI), can lead to chronic infections. Not only the inner parts of the implant but also the externally located headpiece might be associated with prolonged superficial skin eczema resulting in the inability of wearing the headpiece. In this study, the surface of three CI headpieces from different manufacturers were examined for bacterial biofilm formation.

Materials and methods

Two bacterial species associated with implant-related infections were tested: Pseudomonas aeruginosa (ATCC9027) and Staphylococcus aureus (ATCC6538). Biofilms were formed over 24 h in tryptic soy broth at 36 °C. Biofilm formation was detected in form of biomass measurement by crystal violet staining. CI headpiece dummies of three manufacturers were used.

Results

Both tested bacterial species formed biofilms on the examined CI headpiece-surfaces in a species-dependent manner with higher biofilm formation of P. aeruginosa. For both, S. aureus and P. aeruginosa, biofilm formation on the CI components was comparable to a polystyrene control surface. Between the three manufacturers, no significant difference in biofilm formation was found.

Discussion

The tested bacteria displayed biofilm formation on the CI headpieces in a species-specific manner with higher amount of biofilm formed by P. aeruginosa. The biofilm formation was comparable between the manufacturers. In this study, an enhanced biofilm formation on CI headpieces could not be demonstrated. These in vitro tests suggest a minor role of bacterial biofilm on the CI headpiece in skin infections under the CI headpiece.