Introduction <p>Computer-assisted surgery (CAS) in total knee arthroplasty (TKA) has seen growing adoption, with over 30% of primary TKA procedures in Japan using CAS as of 2021. Augmented reality (AR) navigation has emerged as a novel alternative, providing real-time visualization and high-precision guidance during surgery. However, standardized protocols, long-term outcomes, and quantitative accuracy data remain limited. This study aimed to evaluate the intraoperative bone resection accuracy achieved with an AR-based navigation system during TKA.</p> Methods <p>We retrospectively analyzed 100 knees that underwent primary TKA using the NextAR navigation system (Medacta International SA, Switzerland). Bone resection accuracy was assessed by comparing intraoperative cut angles and thicknesses to preoperative plans. Evaluated parameters included femoral coronal, sagittal, and rotational angles; distal and posterior resection thicknesses (medial/lateral); tibial coronal, posterior slope, and rotational angles; tibial resection thicknesses (medial/lateral); and implant size concordance.</p> Results <p>Mean absolute differences from preoperative plans were as follows: femoral coronal angle 0.46° ± 0.48°, sagittal angle 0.72° ± 0.61°, and rotational angle 0.30° ± 0.33°; distal cut thickness 0.56 ± 0.46&#xa0;mm (medial) and 0.54 ± 0.49&#xa0;mm (lateral); posterior condyle cut thickness 0.23 ± 0.35&#xa0;mm (medial) and 0.35 ± 0.44&#xa0;mm (lateral). Tibial parameters showed coronal error of 0.64° ± 0.56°, posterior slope 0.66° ± 0.55°, and rotation 0.25° ± 0.60°; resection thickness 0.73 ± 0.69&#xa0;mm (medial) and 0.79 ± 0.75&#xa0;mm (lateral). Implant size matched 100% with preoperative planning.</p> Conclusion <p>The AR-based navigation system enabled precise three-dimensional intraoperative assessment and execution of bone resections in TKA. All angular and thickness errors remained within 1° and 1&#xa0;mm, respectively, confirming high reproducibility and precision across multiple planes.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Intraoperative evaluation of bone resection accuracy in total knee arthroplasty using an augmented reality-based navigation system

  • Atsushi Sato,
  • Masataka Ota,
  • Toshiharu Miyazawa,
  • Misako Takizawa,
  • Reo Nagasaka,
  • Marika Mukunoki,
  • Kanako Izukashi,
  • Jun Oike,
  • Takayuki Okumo,
  • Saki Yagura,
  • Takayuki Koya,
  • Koji Kanzaki

摘要

Introduction

Computer-assisted surgery (CAS) in total knee arthroplasty (TKA) has seen growing adoption, with over 30% of primary TKA procedures in Japan using CAS as of 2021. Augmented reality (AR) navigation has emerged as a novel alternative, providing real-time visualization and high-precision guidance during surgery. However, standardized protocols, long-term outcomes, and quantitative accuracy data remain limited. This study aimed to evaluate the intraoperative bone resection accuracy achieved with an AR-based navigation system during TKA.

Methods

We retrospectively analyzed 100 knees that underwent primary TKA using the NextAR navigation system (Medacta International SA, Switzerland). Bone resection accuracy was assessed by comparing intraoperative cut angles and thicknesses to preoperative plans. Evaluated parameters included femoral coronal, sagittal, and rotational angles; distal and posterior resection thicknesses (medial/lateral); tibial coronal, posterior slope, and rotational angles; tibial resection thicknesses (medial/lateral); and implant size concordance.

Results

Mean absolute differences from preoperative plans were as follows: femoral coronal angle 0.46° ± 0.48°, sagittal angle 0.72° ± 0.61°, and rotational angle 0.30° ± 0.33°; distal cut thickness 0.56 ± 0.46 mm (medial) and 0.54 ± 0.49 mm (lateral); posterior condyle cut thickness 0.23 ± 0.35 mm (medial) and 0.35 ± 0.44 mm (lateral). Tibial parameters showed coronal error of 0.64° ± 0.56°, posterior slope 0.66° ± 0.55°, and rotation 0.25° ± 0.60°; resection thickness 0.73 ± 0.69 mm (medial) and 0.79 ± 0.75 mm (lateral). Implant size matched 100% with preoperative planning.

Conclusion

The AR-based navigation system enabled precise three-dimensional intraoperative assessment and execution of bone resections in TKA. All angular and thickness errors remained within 1° and 1 mm, respectively, confirming high reproducibility and precision across multiple planes.