错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Intraoperative cone-beam computed tomography for catheter placement verification in pediatric hydrocephalus: technical note

  • Matthias Krause,
  • Jasmina Lagumdzija,
  • Simon Enzinger,
  • Jörn Wittig,
  • Alexander Gaggl,
  • Roman P. Metzger,
  • Christoph J. Griessenauer

摘要

Ventriculoperitoneal (VP) shunt placement, essential for managing hydrocephalus, often risks catheter malpositioning, especially in patients with small ventricles. We present a novel technique combining neuronavigation with intraoperative cone-beam computed tomography using the BrainLab system and Loop-X mobile imaging unit. This approach enables real-time verification of catheter placement by integrating preoperative MRI data with intraoperative CT imaging. In a 12-year-old boy with therapy-refractory idiopathic intracranial hypertension, neuronavigation was guided by the BrainLab Skull Fix and Cushing canula, ensuring precise catheter insertion into the right frontal horn. Post-placement, Loop-X facilitated immediate verification of the catheter’s trajectory and positioning, corroborated by postoperative MRI. This technique demonstrated high precision and minimized radiation exposure, emphasizing its utility in reducing revision rates due to suboptimal catheter placement.