<p>We propose a method to perform accurate temperature measurements using laser-induced fluorescence (LIF). We use a sCMOS color camera and a two-dye solution consisting of RuPhen and fluorescein, excited at 450 <i>nm</i>. By varying the relative concentration of the dyes, we can tune the temperature sensitivity of the color channels. This enables a robust laser power correction, reducing the effects of experimental noise compared to the convectional ratiometric approach. Furthermore, the overall temperature sensitivity is only slightly lower compared to that of the temperature-sensitive dye, which is not the case if a ratiometric analysis is performed. We demonstrate the capabilities of our method using a benchtop setup with precisely controlled temperatures. Our error analysis shows that an accuracy better than 0.5 <InlineEquation ID="IEq1"> <EquationSource Format="TEX">\(^\circ C\)</EquationSource> <EquationSource Format="MATHML"><math> <mrow> <mmultiscripts> <mrow /> <mrow /> <mo>∘</mo> </mmultiscripts> <mi>C</mi> </mrow> </math></EquationSource> </InlineEquation> can be achieved. The correction method can be applied to other fluorescence measurement techniques, including pressure-sensitive paints (PSPs).</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Explicit illumination correction for error reduction in two-dye LIF thermometry

  • Claudio Mucignat,
  • Thomas Rösgen,
  • Ivan Lunati

摘要

We propose a method to perform accurate temperature measurements using laser-induced fluorescence (LIF). We use a sCMOS color camera and a two-dye solution consisting of RuPhen and fluorescein, excited at 450 nm. By varying the relative concentration of the dyes, we can tune the temperature sensitivity of the color channels. This enables a robust laser power correction, reducing the effects of experimental noise compared to the convectional ratiometric approach. Furthermore, the overall temperature sensitivity is only slightly lower compared to that of the temperature-sensitive dye, which is not the case if a ratiometric analysis is performed. We demonstrate the capabilities of our method using a benchtop setup with precisely controlled temperatures. Our error analysis shows that an accuracy better than 0.5 \(^\circ C\) C can be achieved. The correction method can be applied to other fluorescence measurement techniques, including pressure-sensitive paints (PSPs).