EndoArt®: Ergebnisse bei Patienten mit Glaukom-Drainage-Implantaten

  • Johanna Wiedemann,
  • Mert Mestanoglu,
  • Anne Rekate,
  • Caroline Gietzelt,
  • Claus Cursiefen,
  • Björn Bachmann

摘要

Background

EndoArt® (Eye Yon Medical, Ness Ziona, Israel) is a novel artificial corneal inner layer transplant and an innovative treatment alternative for patients at high risk for graft failure after posterior lamellar corneal transplantation (EK).

Aim of the study

We present the initial results of the EndoArt® implantation in patients with glaucoma drainage devices (GDD).

Patients and surgical procedure

In this study 12 eyes with GDD were retrospectively evaluated. All had a high risk of transplant rejection in cases of EK (previous other glaucoma surgery in addition to GDD, n = 8, condition following Descemet membrane endothelial keratoplasty, DMEK, n = 9, uveitis, n = 2, or synechiae, n = 2). The EndoArt® was secured with a gas bubble and one to three holding sutures. The preoperative and postoperative best spectacle-corrected visual acuity (BSCVA) and central corneal thickness (CCT) were determined. The need for additional gas injections (rebubbling) was analyzed.

Results

Octafluoropropane (C3F8) 12% was used in nine patients and sulfur hexafluoride (SF6) 20% in three patients as anterior chamber tamponade to ensure adhesion of the EndoArt®. At least one rebubbling was necessary in four eyes. The preoperative BSCVA was 1.6 (± 0.5) logMAR and significantly improved to 1.1 (± 0.6) logMAR after 12 weeks (p = 0.028). The preoperative CCT was 719 µm ± 145.7, which significantly decreased to 622.4 µm ± 174.9 after six weeks (p = 0.004) and to 591.7 µm ± 190.8 after 12 weeks (p = 0.096). In one eye, the prosthesis was explanted due to a suspected fibrotic remodelling of the cornea following multiple previous DMEKs.

Discussion

EndoArt® led to a significant improvement in BSCVA and reduction in CCT; however, there was a high rebubbling rate in patients with GDD. In high-risk patients, the use of EndoArt® may be advantageous in avoiding graft failure after EK.