Purpose <p>Intrarenal pressure monitoring is emerging as an important tool in flexible ureteroscopy. Novel non-invasive intrarenal pressure monitoring devices are introduced. We evaluated the pressure readings simultaneously across four novel non-invasive intrarenal pressure monitoring devices against the urodynamic system.</p> Materials and methods <p>Two IRP monitoring single-use flexible ureteroscopy: LithoVue Elite<sup>™</sup> (Boston Scientific Corporation, Marlborough, MA, USA); ZebraScope<sup>™</sup> (Happiness Works Medical Instruments Co., Ltd, Anhui, China); two intelligent intrarenal system integrated with irrigation and suction via flexible and navigable suction ureteral access sheath: i-MIMERsys<sup>™</sup> (ZSR Biomedical Technology Co., Ltd, Dongguan, China); Tidor<sup>™</sup> (YiGao Medical Technology Co., Ltd, Zhejiang, China) were tested against MMS/Laborie Nexam Pro Urodynamic System (Laborie Medical Technologies Corporation, Portsmouth, USA).</p> Results <p>The degree of agreement between measurements from urodynamic system and the experimented device was demonstrated by Bland–Altman plots and intraclass correlation coefficients (ICCs). Compared to the urodynamic system, the average difference ± standard deviation of LithoVue Elite™ was 0.26 ± 1.75mmHg; ZebraScope™ 3.22 ± 1.86mHg, i-MIMERsys™ 0.96 ± 2.26mmhg; Tidor™ 1.09 ± 2.57mmHg. Pressure measured by LithoVue Elite™, ZebraScope™, i-MIMERsys™ and Tidor™ to urodynamic system showed excellent ICCs of 0.975 (<i>p</i> &lt; 0.001), 0.915(<i>p</i> = 0.016), 0.955(<i>p</i> &lt; 0.001) and 0.937(<i>p</i> &lt; 0.001) respectively.</p> Conclusion <p>Pressure recordings from four novel non-invasive intrarenal monitoring devices, namely flexible ureteroscopes—LithoVue Elite<sup>™</sup>, ZebraScope<sup>™</sup>; and intelligent IRP monitoring systems via flexible and navigable suction ureteral access sheaths — i-MIMERsys<sup>™</sup>, Tidor<sup>™</sup>. were comparable to that of urodynamic system. This in-vitro study verifies that such non-invasive devices provide reliable data for monitoring intrarenal pressure changes.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Novel non-invasive intrarenal pressure monitoring devices in flexible ureteroscopy: an in-vitro comparative study

  • Steffi Kar Kei Yuen,
  • Zixiao Chen,
  • Skyler Yuen,
  • Alex Qinyang Liu,
  • Chi-Ho Leung,
  • Ivan Ching Ho Ko,
  • Chi Kwok Chan,
  • Bhaskar Somani,
  • Thomas Herrmann,
  • Helene Jung,
  • Theodoros Tokas,
  • Stefanie Croghan,
  • Olivier Traxer,
  • Peter Ka Fung Chiu,
  • Ben Chew,
  • Jeremy Yuen Chun Teoh,
  • Vineet Gauhar,
  • Zheng Li,
  • Chi Fai Ng

摘要

Purpose

Intrarenal pressure monitoring is emerging as an important tool in flexible ureteroscopy. Novel non-invasive intrarenal pressure monitoring devices are introduced. We evaluated the pressure readings simultaneously across four novel non-invasive intrarenal pressure monitoring devices against the urodynamic system.

Materials and methods

Two IRP monitoring single-use flexible ureteroscopy: LithoVue Elite (Boston Scientific Corporation, Marlborough, MA, USA); ZebraScope (Happiness Works Medical Instruments Co., Ltd, Anhui, China); two intelligent intrarenal system integrated with irrigation and suction via flexible and navigable suction ureteral access sheath: i-MIMERsys (ZSR Biomedical Technology Co., Ltd, Dongguan, China); Tidor (YiGao Medical Technology Co., Ltd, Zhejiang, China) were tested against MMS/Laborie Nexam Pro Urodynamic System (Laborie Medical Technologies Corporation, Portsmouth, USA).

Results

The degree of agreement between measurements from urodynamic system and the experimented device was demonstrated by Bland–Altman plots and intraclass correlation coefficients (ICCs). Compared to the urodynamic system, the average difference ± standard deviation of LithoVue Elite™ was 0.26 ± 1.75mmHg; ZebraScope™ 3.22 ± 1.86mHg, i-MIMERsys™ 0.96 ± 2.26mmhg; Tidor™ 1.09 ± 2.57mmHg. Pressure measured by LithoVue Elite™, ZebraScope™, i-MIMERsys™ and Tidor™ to urodynamic system showed excellent ICCs of 0.975 (p < 0.001), 0.915(p = 0.016), 0.955(p < 0.001) and 0.937(p < 0.001) respectively.

Conclusion

Pressure recordings from four novel non-invasive intrarenal monitoring devices, namely flexible ureteroscopes—LithoVue Elite, ZebraScope; and intelligent IRP monitoring systems via flexible and navigable suction ureteral access sheaths — i-MIMERsys, Tidor. were comparable to that of urodynamic system. This in-vitro study verifies that such non-invasive devices provide reliable data for monitoring intrarenal pressure changes.