An in vitro study of irrigation rate, operator duty cycle and intrarenal temperature in superpulse fiber thulium laser lithotripsy
摘要
The present study aimed to examine the relationship between irrigation velocity, operator duty cycle (ODC), and intrarenal temperature during retrograde intrarenal surgery with a superpulse fiber thulium laser.
MethodsPlace the stones into the fresh isolated porcine kidneys, use puncture needle to place the temperature probe 2 mm around the stones, and place the pressure probes in the upper calyx, lower calyx, and renal pelvis. Place the entire setup in a 37 °C constant temperature water bath to simulate the human body environment. The laser power varies between 10 and 30 W, and the irrigation speed is 10-30 ml/min. Additionally, at a laser power of 20 W and an irrigation speed of 10 ml/min, different On-Duty Cycles (ODC) are set. Monitor the changes in temperature and pressure.
ResultsA direct proportionality of temperature in the kidney to the rate of irrigation has been reported between 10 W and 30 W laser powers. The percentage ratio of the rate of irrigation and power in the laser is 1:1, which can keep the temperature in the kidney at a safe level. At a laser power of 20 W and irrigation of 10 ml/min, the temperature inside the kidney increases sharply with the increase in ODC. By decreasing the ratio of ODC, the increase of temperature inside the kidney can be brought to a great reduction.
ConclusionMaintaining a 1:1 ratio between laser power and irrigation speed can effectively prevent thermal damage or injury to kidney tissue.Additionally, by adjusting the On-Duty Cycle (ODC) ratio, the intrarenal temperature can also be reduced.