Temperature profile during endourological laser activation: introducing the thermal safety distance concept

  • Eugenio Ventimiglia,
  • Daniele Robesti,
  • Etienne Xavier Keller,
  • Christian Corsini,
  • Andrea Folcia,
  • Margherita Fantin,
  • Luigi Candela,
  • Edoardo Pozzi,
  • Alba Sierra,
  • Amelia Pietropaolo,
  • Bhaskar K. Somani,
  • Frederic Panthier,
  • Felipe Pauchard,
  • Ioannis Kartalas Goumas,
  • Luca Villa,
  • Francesco Montorsi,
  • Olivier Traxer,
  • Andrea Salonia,
  • Paola Saccomandi

摘要

Purpose

To examine temporal-spatial distribution of heat generated upon laser activation in a bench model of renal calyx. To establish reference values for a safety distance between the laser fiber and healthy tissue during laser lithotripsy.

Methods

We developed an in-vitro experimental setup employing a glass pipette and laser activation under various intra-operative parameters, such as power and presence of irrigation. A thermal camera was used to monitor both temporal and spatial temperature changes during uninterrupted 60-second laser activation. We computed the thermal dose according to Sapareto and Dewey’s formula at different distances from the laser fiber tip, in order to determine a safety distance.

Results

A positive correlation was observed between average power and the highest recorded temperature (Spearman’s coefficient 0.94, p < 0.001). Irrigation was found to reduce the highest recorded temperature, with a maximum average reduction of 9.4 °C at 40 W (p = 0.002). A positive correlation existed between average power and safety distance values (Spearman’s coefficient 0.86, p = 0.001). A thermal dose indicative of tissue damage was observed at 20 W without irrigation (safety distance 0.93±0.11 mm). While at 40 W, irrigation led to slight reduction in mean safety distance (4.47±0.85 vs. 5.22±0.09 mm, p = 0.08).

Conclusions

Laser settings with an average power greater than 10 W deliver a thermal dose indicative of tissue damage, which increases with higher average power values. According to safety distance values from this study, a maximum of 10 W should be used in the ureter, and a maximum of 20 W should be used in kidney in presence of irrigation.