Illumination matters Part II: advanced comparative analysis of flexible ureteroscopes in a kidney model by PEARLS

  • Jia-Lun Kwok,
  • Frédéric Panthier,
  • Vincent De Coninck,
  • Eugenio Ventimiglia,
  • Yazeed Barghouthy,
  • Alexandre Danilovic,
  • Niamh Smyth,
  • Jan Brachlow,
  • Florian Alexander Schmid,
  • Cédric Poyet,
  • Daniel Eberli,
  • Olivier Traxer,
  • Etienne Xavier Keller

摘要

Purpose

The aim of the study was to evaluate illumination properties in an in-vitro kidney calyx model in saline.

Design and methods

We evaluated a series of contemporary flexible ureteroscopes including the Storz Flex-Xc and Flex-X2s, Olympus V3 and P7, Pusen 7.5F and 9.2F, as well as OTU WiScope using a 3D-printed closed pink kidney calyx model, submerged in saline. A spectrometer was used for illuminance and color temperature measurements at different openings located at center (direct light), 45° (direct and indirect light) and 90°(indirect light) to the axis of the scope.

Results

Maximum illuminance was at the center opening for all scopes (range: 284 to 12,058 lx at 50% brightness and 454 to 11,871 lx at 100% brightness settings). The scope with the highest center illuminance (Flex-Xc) was 26 times superior to the scope with the lowest illuminance (Pusen 7.5Fr) at 100% brightness setting. For each scope, there was a peripheral illuminance drop ranging from − 43 to − 92% at 50% brightness and − 43% to − 88% at 100% brightness settings, respectively (all p < 0.01). Highest drop was for the P7 and the Pusen 9.2F. All scopes had illuminance skew, except the V3. All scopes had a warm color temperature.

Conclusion

Illumination properties vary between ureteroscopes in an enclosed cavity in saline, and differs at center vs 45° and 90° positions within scopes. Peripheral illuminance drop can be as high as − 92%, which is undesirable. This may affect the choice of ureteroscope and light brightness settings used in surgery by urologists.