Comparison of hepatic iron quantification by MRI between 1.5 and 3 T, using different methods and conversion formulas
摘要
This study aimed to compare liver iron concentration (LIC) estimates obtained from 1.5-T and 3-T MRI systems, using multiple quantification methods.
Materials and methodsSeventy-eight participants from two centers were prospectively included and underwent liver MRI scans at both 1.5 T and 3 T. Two multi-echo gradient recalled echo (ME-GRE) sequences—one vendor-specific and one non-specific—were used to quantify liver fat and iron. Proton density fat fraction (PDFF) and R2* were obtained from vendor sequences via map reconstruction, and from non-specific sequences using MRQuantif software. LIC was derived from R2* using European or American calibrations, but also from signal intensity ratio (SIR) method at 1.5 T. Statistical analysis included linear correlation regression and Bland–Altman plots to evaluate agreement between LIC values across field strengths and methods.
ResultsFor LIC calculated from R2*, both sequences demonstrated a strong correlation between 1.5-T and 3-T results, with R = 0.84 and 0.99 for the vendor and MRQuantif methods, respectively. The LIC bias between field strengths remained low (± 7% maximum) when using the same calibration group or when comparing, at 1.5 T, the SIR method and European calibrations. However, this bias increased to 26% when comparing LIC values obtained from American and European calibrations.
ConclusionLIC values are highly comparable when similar calibration methods are applied, but notable bias arises when comparing results from different calibration groups. For clinical convenience, a rapid and intermediate LIC (in µmol/g) can be estimated by dividing R2*values by 2 at 1.5 T or by 4 at 3 T.
Key Points