Detection of lumbar disc herniation and degeneration using photon-counting detector CT with virtual non-calcium imaging
摘要
This study aimed to compare the diagnostic performance of virtual non-calcium (VNCa) imaging using photon-counting detector CT (PCD-CT) with MRI in assessing lumbar disc herniation (LDH) and intervertebral disc degeneration (IDD).
Materials and methodsPatients with low back pain underwent PCD-CT and MRI within 2 h. Quantitative analysis included VNCa-derived CT values and T2 relaxation times in manually placed regions of interest. Qualitative assessment focused on diagnostic accuracy for LDH and root impingement using grayscale and color-coded VNCa images.
ResultsA total of 68 participants (51.8% female; mean age, 48 ± 14 years) were included. A significant negative correlation was observed between VNCa CT and MRI T2 values (r = −0.705, p < 0.001), strongest in the nucleus pulposus. Compared with grayscale CT, color-coded VNCa imaging significantly improved sensitivity (81–94%), specificity (84–95%), and accuracy (84–95%) for detecting LDH (all comparisons, p < 0.001); and improved sensitivity (82–94%), specificity (96–97%), and accuracy (93–96%) for diagnosing root impingement (all comparisons, p < 0.05).
ConclusionsPCD-CT with VNCa imaging enables reliable quantification of disc water content and significantly improves diagnostic accuracy for LDH and root impingement. It offers a valuable alternative to MRI, especially in clinical settings where MRI is limited or unavailable.
Key Points