Performance of a screening-trained DL model for pulmonary nodule malignancy estimation of incidental clinical nodules
摘要
To test the performance of a DL model developed and validated for screen-detected pulmonary nodules on incidental nodules detected in a clinical setting.
Materials and methodsA retrospective dataset of incidental pulmonary nodules sized 5–15 mm was collected, and a subset of size-matched solid nodules was selected. The performance of the DL model was compared to the Brock model. AUCs with 95% CIs were compared using the DeLong method. Sensitivity and specificity were determined at various thresholds, using a 10% threshold for the Brock model as reference. The model’s calibration was visually assessed.
ResultsThe dataset included 49 malignant and 359 benign solid or part-solid nodules, and the size-matched dataset included 47 malignant and 47 benign solid nodules. In the complete dataset, AUCs [95% CI] were 0.89 [0.85, 0.93] for the DL model and 0.86 [0.81, 0.92] for the Brock model (p = 0.27). In the size-matched subset, AUCs of the DL and Brock models were 0.78 [0.69, 0.88] and 0.58 [0.46, 0.69] (p < 0.01), respectively. At a 10% threshold, the Brock model had a sensitivity of 0.49 [0.35, 0.63] and a specificity of 0.92 [0.89, 0.94]. At a threshold of 17%, the DL model matched the specificity of the Brock model at the 10% threshold, but had a higher sensitivity (0.57 [0.43, 0.71]). Calibration analysis revealed that the DL model overestimated the malignancy probability.
ConclusionThe DL model demonstrated good discriminatory performance in a dataset of incidental nodules and outperformed the Brock model, but may need recalibration for clinical practice.
Key Points