Objectives <p>To compare image quality and iodine attenuation intra-individually in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium (CM) volume.</p> Materials and methods <p>A prospectively acquired patient cohort between 04/2021 and 11/2023 was retrospectively screened if patients had the following combination of portal venous phase thoracoabdominal CT scans: (a) PCD-CT with 120 mL CM volume (PCD-CT<sub>120 mL</sub>), (b) PCD-CT with 100 mL CM volume (PCD-CT<sub>100 mL</sub>), and (c) prior energy-integrating detector CT (EID-CT) with 120 mL CM volume. On PCD-CT, virtual monoenergetic image (VMI) reconstructions at 70 keV were applied for both groups as well as additional VMI at 60 keV for PCD‑CT<sub>100 mL</sub>. Quantitative analyses including signal-to-noise (SNR) and contrast-to-noise ratios (CNR) and qualitative analyses were performed using a mixed linear effects model.</p> Results <p>The final study cohort comprised 49 patients (mean age 67 [31–86] years, 12 female). Comparison to EID-CT was available in 33 patients. In standard 70 keV VMI reconstructions, PCD-CT<sub>100 mL</sub> was non-inferior to PCD-CT<sub>120 mL</sub> as well as to EID-CT<sub>120 mL</sub> for CNR in abdominal organs (all <i>p</i> &gt; 0.050). The mixed linear effects model revealed significant differences between contrast volume groups for both contrast enhancement and image quality ratings. PCD-CT<sub>100 mL/70 keV</sub> demonstrated the smallest deviation from optimal contrast enhancement (−0.306, <i>p</i> &lt; 0.001).</p> Conclusion <p>In portal venous phase thoracoabdominal PCD-CT, a nearly 17% reduction in CM was achievable while maintaining subjective and objective image quality compared to prior higher CM volume PCD-CT scans within the same patients and still surpassing image quality of previous exams on an EID-CT system.</p> Key Points <p><Emphasis Type="BoldItalic">Question</Emphasis> <i>How do image quality and iodine attenuation intra-individually compare in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium volume.</i></p> <p><Emphasis Type="BoldItalic">Findings</Emphasis> <i>PCD-CT scans exhibit superior quantitative and qualitative image quality compared to energy-integrating detector-CT acquisitions and are not negatively affected by contrast volume reductions up to 17%.</i></p> <p><Emphasis Type="BoldItalic">Clinical relevance</Emphasis> <i>This study provides further evidence that PCD-CT enables a considerable reduction in iodine dose for portal venous phase acquisition, benefiting both patients and healthcare system costs.</i></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Potential of photon-counting detector CT technology for contrast medium reduction in portal venous phase thoracoabdominal CT

  • Daniel Popp,
  • Martin Siedlecki,
  • Lena Friedrich,
  • Mark Haerting,
  • Christian Scheurig-Muenkler,
  • Florian Schwarz,
  • Thomas Kroencke,
  • Stefanie Bette,
  • Josua A. Decker

摘要

Objectives

To compare image quality and iodine attenuation intra-individually in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium (CM) volume.

Materials and methods

A prospectively acquired patient cohort between 04/2021 and 11/2023 was retrospectively screened if patients had the following combination of portal venous phase thoracoabdominal CT scans: (a) PCD-CT with 120 mL CM volume (PCD-CT120 mL), (b) PCD-CT with 100 mL CM volume (PCD-CT100 mL), and (c) prior energy-integrating detector CT (EID-CT) with 120 mL CM volume. On PCD-CT, virtual monoenergetic image (VMI) reconstructions at 70 keV were applied for both groups as well as additional VMI at 60 keV for PCD‑CT100 mL. Quantitative analyses including signal-to-noise (SNR) and contrast-to-noise ratios (CNR) and qualitative analyses were performed using a mixed linear effects model.

Results

The final study cohort comprised 49 patients (mean age 67 [31–86] years, 12 female). Comparison to EID-CT was available in 33 patients. In standard 70 keV VMI reconstructions, PCD-CT100 mL was non-inferior to PCD-CT120 mL as well as to EID-CT120 mL for CNR in abdominal organs (all p > 0.050). The mixed linear effects model revealed significant differences between contrast volume groups for both contrast enhancement and image quality ratings. PCD-CT100 mL/70 keV demonstrated the smallest deviation from optimal contrast enhancement (−0.306, p < 0.001).

Conclusion

In portal venous phase thoracoabdominal PCD-CT, a nearly 17% reduction in CM was achievable while maintaining subjective and objective image quality compared to prior higher CM volume PCD-CT scans within the same patients and still surpassing image quality of previous exams on an EID-CT system.

Key Points

Question How do image quality and iodine attenuation intra-individually compare in portal venous phase photon-counting detector CT (PCD-CT) scans using protocols with different contrast medium volume.

Findings PCD-CT scans exhibit superior quantitative and qualitative image quality compared to energy-integrating detector-CT acquisitions and are not negatively affected by contrast volume reductions up to 17%.

Clinical relevance This study provides further evidence that PCD-CT enables a considerable reduction in iodine dose for portal venous phase acquisition, benefiting both patients and healthcare system costs.