Objectives <p>To examine CT radiation dose variation in PET-CT and understand how often providers use diagnostic vs lower doses for attenuation correction and anatomic localization (AC/AL).</p> Methods <p>Retrospective, multi-centered study of PET-CT exams from 2010 to 2021. Exams were categorized as body or brain. Radiation dose was quantified using dose length product (DLP), adjusted for patient size (DLP<sub>S-ADJ</sub>), and for size and scan length (DLP<sub>SL-ADJ</sub>). DLP<sub>S-ADJ</sub> variation was assessed by body region and facility. To ascertain whether sites use AC/AL or diagnostic doses, we compared each facility’s DLP<sub>S-ADJ</sub> in CTs from PET-CT vs diagnostic CT exams of analogous body regions (not associated with PET), controlling for scan length using DLP<sub>SL-ADJ</sub>. Lastly, we categorized exams as likely diagnostic if they used multiple phases and/or contrast to compare dose and frequency with likely non-diagnostic exams.</p> Results <p>Sixty-four thousand two hundred ten exams included 93% body, 7% brain. Doses were higher and more variable in the body than brain exams (adult mean DLP<sub>S-ADJ</sub> = 1004 vs 341 mGy-cm, respectively). For body exams, DLP<sub>S-ADJ</sub> was higher in older children than adults, and there was wide inter-facility variation (median DLP<sub>S-ADJ</sub> range = 245–2391 mGy-cm). Most facilities used higher CT doses in body PET-CT than in diagnostic CT exams even controlling for scan length. Fifty-three percent of adult and 76% of child body exams were likely diagnostic based on the use of diagnostic techniques.</p> Conclusion <p>While diagnostic CT is sometimes indicated for PET-CT, body exams generally do not use AC/AL protocols. Doses were higher than previously reported, higher than analogous diagnostic CT exams, and higher in older children than adults when size adjusted.</p> Key Points <p><Emphasis Type="BoldItalic">Question&#xa0;</Emphasis><i>How does CT radiation dose vary in PET-CT and how often do providers use diagnostic vs lower doses for AC/AL</i>?</p> <p><Emphasis Type="BoldItalic">Findings</Emphasis> <i>Most facilities did not use AC/AL protocols for body PET-CT, and CT effective doses were higher than previously reported</i>.</p> <p><Emphasis Type="BoldItalic">Clinical relevance</Emphasis> <i>The considerable inter-facility variation observed suggests ample opportunity to reduce ionizing radiation doses for CT in body PET-CT by adopting low-dose (AC/AL) protocols and other dose reduction techniques</i>.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Use of diagnostic vs low-dose computed tomography in positron emission tomography-CT examinations

  • Matthew Malekhedayat,
  • Carly Stewart,
  • Philip W. Chu,
  • Yifei Wang,
  • Nima Kasraie,
  • Benjamin Franc,
  • Rebecca Smith-Bindman

摘要

Objectives

To examine CT radiation dose variation in PET-CT and understand how often providers use diagnostic vs lower doses for attenuation correction and anatomic localization (AC/AL).

Methods

Retrospective, multi-centered study of PET-CT exams from 2010 to 2021. Exams were categorized as body or brain. Radiation dose was quantified using dose length product (DLP), adjusted for patient size (DLPS-ADJ), and for size and scan length (DLPSL-ADJ). DLPS-ADJ variation was assessed by body region and facility. To ascertain whether sites use AC/AL or diagnostic doses, we compared each facility’s DLPS-ADJ in CTs from PET-CT vs diagnostic CT exams of analogous body regions (not associated with PET), controlling for scan length using DLPSL-ADJ. Lastly, we categorized exams as likely diagnostic if they used multiple phases and/or contrast to compare dose and frequency with likely non-diagnostic exams.

Results

Sixty-four thousand two hundred ten exams included 93% body, 7% brain. Doses were higher and more variable in the body than brain exams (adult mean DLPS-ADJ = 1004 vs 341 mGy-cm, respectively). For body exams, DLPS-ADJ was higher in older children than adults, and there was wide inter-facility variation (median DLPS-ADJ range = 245–2391 mGy-cm). Most facilities used higher CT doses in body PET-CT than in diagnostic CT exams even controlling for scan length. Fifty-three percent of adult and 76% of child body exams were likely diagnostic based on the use of diagnostic techniques.

Conclusion

While diagnostic CT is sometimes indicated for PET-CT, body exams generally do not use AC/AL protocols. Doses were higher than previously reported, higher than analogous diagnostic CT exams, and higher in older children than adults when size adjusted.

Key Points

Question How does CT radiation dose vary in PET-CT and how often do providers use diagnostic vs lower doses for AC/AL?

Findings Most facilities did not use AC/AL protocols for body PET-CT, and CT effective doses were higher than previously reported.

Clinical relevance The considerable inter-facility variation observed suggests ample opportunity to reduce ionizing radiation doses for CT in body PET-CT by adopting low-dose (AC/AL) protocols and other dose reduction techniques.