Implementing an AI algorithm in the clinical setting: a case study for the accuracy paradox
摘要
We report our experience implementing an algorithm for the detection of large vessel occlusion (LVO) for suspected stroke in the emergency setting, including its performance, and offer an explanation as to why it was poorly received by radiologists.
Materials and methodsAn algorithm was deployed in the emergency room at a single tertiary care hospital for the detection of LVO on CT angiography (CTA) between September 1st–27th, 2021. A retrospective analysis of the algorithm’s accuracy was performed.
ResultsDuring the study period, 48 patients underwent CTA examination in the emergency department to evaluate for emergent LVO, with 2 positive cases (60.3 years ± 18.2; 32 women). The LVO algorithm demonstrated a sensitivity and specificity of 100% and 92%, respectively. While the sensitivity of the algorithm at our institution was even higher than the manufacturer’s reported values, the false discovery rate was 67%, leading to the perception that the algorithm was inaccurate. In addition, the positive predictive value at our institution was 33% compared with the manufacturer’s reported values of 95–98%. This disparity can be attributed to differences in disease prevalence of 4.1% at our institution compared with 45.0–62.2% from the manufacturer’s reported values.
ConclusionDespite the LVO algorithm’s accuracy performing as advertised, it was perceived as inaccurate due to more false positives than anticipated and was removed from clinical practice. This was likely due to a cognitive bias called the accuracy paradox. To mitigate the accuracy paradox, radiologists should be presented with metrics based on a disease prevalence similar to their practice when evaluating and utilizing artificial intelligence tools.
Key Points