3D full-dose brain-PET volume recovery from low-dose data through deep learning: quantitative assessment and clinical evaluation
摘要
Low-dose (LD) PET imaging would lead to reduced image quality and diagnostic efficacy. We propose a deep learning (DL) method to reduce radiotracer dosage for PET studies while maintaining diagnostic quality.
MethodsThis retrospective study was performed on 456 participants respectively scanned by three different PET scanners with two different tracers. A DL method called spatially aware noise reduction network (SANR) was proposed to recover 3D full-dose (FD) PET volumes from LD data. The performance of SANR was compared with a 2D DL method taking regular FD PET volumes as the reference. Wilcoxon signed-rank test was conducted to compare the image quality metrics across different DL denoising methods. For clinical evaluation, two nuclear medicine physicians examined the recovered FD PET volumes using a 5-point grading scheme (5 = excellent) and gave a binary decision (negative or positive) for diagnostic quality assessment.
ResultsStatistically significant differences (p < 0.05) were found in terms of image quality metrics when SANR was compared with the 2D DL method. For clinical evaluation, SANR achieved a lesion detection accuracy of 95.3% (95% CI: 90.1%, 100%), while the reference full-dose PET volumes obtained a lesion detection accuracy of 98.4% (95% CI: 95.4%, 100%). In Alzheimer’s disease diagnosis, both the reference FD PET volumes and the FD PET volumes recovered by SANR exhibited the same accuracy.
ConclusionCompared with reference FD PET, LD PET denoised by the proposed approach significantly reduced radiotracer dosage and showed noninferior diagnostic performance in brain lesion detection and Alzheimer’s disease diagnosis.
Key Points