Photon-counting detector CT provides superior subsolid nodule characterization compared to same-day energy-integrating detector CT
摘要
To investigate the image quality and the performance of photon-counting detector (PCD) CT compared to conventional energy-integrating detector (EID) CT in identifying subsolid nodule (SSN) characteristics.
Materials and methodsParticipants with SSNs who underwent same-day EID CT and PCD CT between October 2023 and April 2024 were prospectively included. The 1.0 mm EID CT images and, subsequently, 1.0 mm, 0.4 mm, and 0.2 mm PCD CT images were reviewed to assess image noise and subjective image quality on a 5-point Likert scale. SSN characteristics, including lobulation, spiculation, pleural retraction, air cavities, intra-nodular vessel signs, internal vascular changes, and heterogeneous solid components, were evaluated. Additionally, a step-by-step observation and comparison method was used to determine the presence of any additional characteristics.
ResultsForty-eight participants (mean age: 56 ± 11 years; 16 males) with 89 SSNs were included. PCD CT significantly reduced radiation dose when using matched scans (1.79 ± 0.39 vs 2.17 ± 0.57 mSv, p < 0.001). Compared to 1.0 mm EID CT, 1.0 mm PCD CT images exhibited significantly lower objective image noise and higher subjective image quality (all p < 0.001). Compared to EID CT, PCD CT demonstrated enhanced visualization of subtle characteristics, except for lobulation, with a 0.4 mm section thickness offering a favorable balance between ultra-high resolution and perceived image quality for radiologists.
ConclusionPCD CT facilitated radiation dose reduction and outperformed conventional EID CT in terms of image quality and visualization of SSN characteristics.
Key Points