Feasibility and Safety of Percutaneous Puncture Guided by a 5G-Based Telerobotic Ultrasound System: An Experimental Study

  • Lanling Yang,
  • Shaobo Duan,
  • Ye Zhang,
  • Liuwei Hao,
  • Shuaiyang Wang,
  • Zhi Zou,
  • Yanshan Hu,
  • Si Chen,
  • Yiwen Hu,
  • Lianzhong Zhang

摘要

Purpose

To evaluate the feasibility and safety of percutaneous puncture guided by a 5th generation mobile communication technology (5G)-based telerobotic ultrasound system in phantom and animal experiments.

Materials and Methods

In the phantom experiment, 10 simulated lesions were punctured, once at each of two angles for each lesion, under the guidance of a telerobotic ultrasound system and ultrasound-guided freehand puncture. Student’s t test was used to compare the two methods in terms of puncture accuracy, total operation duration, and puncture duration. In the animal experiment, under the guidance of the telerobotic ultrasound system, an 18G puncture needle was used to puncture 3 target steel beads in the liver, right kidney, and right gluteal muscle, respectively. The animal experiment had no freehand ultrasound-guided control group. After puncture, a CT scan was performed to verify the position of the puncture needle in relation to the target, and the complications and puncture duration, etc., were recorded.

Results

In the phantom experiment, the mean accuracies of puncture under telerobotic ultrasound guidance and conventional ultrasound guidance were 1.8 ± 0.3 mm and 1.6 ± 0.3 mm (P = 0.09), respectively; therefore, there was no significant difference in the accuracy of the two guide methods. In the animal experiment, the first-attempt puncture success (the needle tip close to the target) rate was 93%. Polypnea occurred during one puncture. No other intraoperative or postoperative complications were observed.

Conclusion

Puncture guided by a 5G-based telerobotic ultrasound system has shown good feasibility and safety in phantom and animal experiments.