Purpose <p> <!--Query ID="Q1" Text="Please check and confirm if the authors and their respective affiliations have been correctly identified. Amend if necessary." Resolved="yes"-->Study of the sagittal accuracy of the ‘Robotic Surgical Assistant’ (ROSA<sup>®</sup>), compared to conventional surgery, regarding the application of the tibial slope (TS). Study of the impact of TS on the range of motion (ROM) and patient-reported outcome measures (PROMS).</p> Methods <p>Inclusion of patients who underwent primary Total Knee Arthroplasty (TKA) between 1/1/2021 and 15/4/2024. Divided into robotic-assisted TKA (RA-TKA) and manual TKA (M-TKA). Measurement of pre- and post-operative TS, using the posterior tibial cortex, on profile knee X-rays. 3° TS applied arbitrarily for both groups. ROM was measured pre-operatively and at three, six and 12 months post-operatively. Patient satisfaction assessed via Knee Injury and Osteoarthritis Outcome Score (KOOS) and Oxford Knee Score (OKS).</p> Results <p>266 patients were included in the study. The M-TKA (110) had a post-operative TS of 3.11° (± 2.12°). 81.21% were within 2° of the target and 92.87% within 3°. The RA-TKA (82) had a post-operative TS of -0.11° ± (1.93°). 36.83% were within 2° of the target and 56.63% within 3°. RA-TKA had a KOOS of 64.43 ± 12.87 and OKS of 33.05 ± 6.01. M-TKA had a KOOS of 64.18 ± 13.11 and OKS of 32.31 ± 5.97. Maximum flexion at 12 months was 118.74° ± 8.19° for M-TKA and 121.88° ± 7.43° for RA-TKA (<i>p</i> = 0.002).</p> Conclusion <p>The application of TS using ROSA<sup>®</sup> was less precise than the conventional method in achieving post-operative TS values as measured on X-rays. However, there were no clinical differences in ROM or PROMS.</p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Sagittal accuracy and functional impact of tibial slope in imageless robotic-assisted Total Knee Arthroplasty

  • Erwan Eggermont,
  • Raphael Janssens,
  • Maarten Ulrix,
  • Jean-François Fils,
  • Jacques Hernigou,
  • Johnatan Everaert,
  • Bruno Baillon

摘要

Purpose

Study of the sagittal accuracy of the ‘Robotic Surgical Assistant’ (ROSA®), compared to conventional surgery, regarding the application of the tibial slope (TS). Study of the impact of TS on the range of motion (ROM) and patient-reported outcome measures (PROMS).

Methods

Inclusion of patients who underwent primary Total Knee Arthroplasty (TKA) between 1/1/2021 and 15/4/2024. Divided into robotic-assisted TKA (RA-TKA) and manual TKA (M-TKA). Measurement of pre- and post-operative TS, using the posterior tibial cortex, on profile knee X-rays. 3° TS applied arbitrarily for both groups. ROM was measured pre-operatively and at three, six and 12 months post-operatively. Patient satisfaction assessed via Knee Injury and Osteoarthritis Outcome Score (KOOS) and Oxford Knee Score (OKS).

Results

266 patients were included in the study. The M-TKA (110) had a post-operative TS of 3.11° (± 2.12°). 81.21% were within 2° of the target and 92.87% within 3°. The RA-TKA (82) had a post-operative TS of -0.11° ± (1.93°). 36.83% were within 2° of the target and 56.63% within 3°. RA-TKA had a KOOS of 64.43 ± 12.87 and OKS of 33.05 ± 6.01. M-TKA had a KOOS of 64.18 ± 13.11 and OKS of 32.31 ± 5.97. Maximum flexion at 12 months was 118.74° ± 8.19° for M-TKA and 121.88° ± 7.43° for RA-TKA (p = 0.002).

Conclusion

The application of TS using ROSA® was less precise than the conventional method in achieving post-operative TS values as measured on X-rays. However, there were no clinical differences in ROM or PROMS.