Feasibility of PET-enabled dual-energy CT imaging: First physical phantom and initial patient study results

  • Yansong Zhu,
  • Siqi Li,
  • Zhaoheng Xie,
  • Edwin K. Leung,
  • Reimund Bayerlein,
  • Negar Omidvari,
  • Yasser G. Abdelhafez,
  • Simon R. Cherry,
  • Jinyi Qi,
  • Ramsey D. Badawi,
  • Benjamin A. Spencer,
  • Guobao Wang

摘要

Purpose

Dual-energy (DE) CT enables material decomposition by using two different x-ray energies and may be combined with PET for improved multimodality imaging. However, this increases radiation dose and may require a hardware upgrade due to the added second x-ray CT scan. The recently proposed PET-enabled DECT method allows dual-energy imaging using a conventional PET/CT scanner without the need to change scanner hardware or increase radiation exposure. Here we demonstrate the first-time physical phantom and patient data evaluation of this method.

Methods

The PET-enabled DECT method reconstructs a gamma-ray CT (gCT) image at 511 keV from the time-of-flight PET data with the maximum-likelihood attenuation and activity (MLAA) approach and then combines this image with the low-energy x-ray CT images to form a dual-energy image pair for material decomposition. To improve the image quality of gCT, a kernel MLAA method was developed using the x-ray CT as a priori information. Here we developed a general open-source implementation for gCT reconstruction and used this implementation for the first real data validation using both physical phantom study and human-subject study. Results from PET-enabled DECT were compared using x-ray DECT as the reference. Further, we applied the PET-enabled DECT method in another patient study to evaluate bone lesions.

Results

Compared to the standard MLAA, results from the kernel MLAA showed significantly improved image quality. PET-enabled DECT with the kernel MLAA was able to generate fractional images that were comparable to the x-ray DECT, with high correlation coefficients for both the phantom study and human subject study (R > 0.99). The application study also indicates that PET-enabled DECT has potential to characterize bone lesions.

Conclusion

Results from this study have demonstrated the feasibility of this PET-enabled method for CT imaging and material decomposition. PET-enabled DECT shows promise to provide comparable results to x-ray DECT.