Background <p>Previous studies have confirmed that fully automated three-dimensional echocardiography (3DE) right ventricular (RV) quantification software can accurately assess adult RV function. However, data on its accuracy in children are scarce.</p> Objective <p>This study aimed to test the accuracy of the software in children using cardiac magnetic resonance (MR) as the gold standard.</p> Materials and methods <p>This study prospectively enrolled 82 children who underwent both echocardiography and cardiac MR within 24&#xa0;h. The RV end-diastolic volume (EDV), end-systolic volume (ESV), and ejection fraction (EF) were obtained using the novel 3DE-RV quantification software and compared with cardiac MR values across different groups.</p> Results <p>The novel 3DE-RV quantification software was feasible in all 82 children (100%). Fully automated analysis was achieved in 35% patients with an analysis time of 8 ± 2&#xa0;s and 100% reproducibility. Manual editing was necessary in the remaining 65% patients. The 3DE-derived RV volumes and EF correlated well with cardiac MR measurements (RVEDV, <i>r</i>=0.93; RVESV, <i>r</i>=0.90; RVEF, <i>r</i>=0.82; all <i>P&#xa0;</i>&lt;0.001). Although the automated approach slightly underestimated RV volumes and overestimated RVEF compared with cardiac MR in the entire cohort, the bias was smaller in children with RVEF ≥ 45%, normal RV size, and good 3DE image quality.</p> Conclusion <p>Fully automated 3DE-RV quantification software provided accurate and completely reproducible results in 35% children without any adjustment. The RV volumes and EF measured using the automated 3DE method correlated well with those from cardiac MR, especially in children with RVEF ≥ 45%, normal RV size, and good 3DE image quality. Therefore, the novel automated 3DE method may achieve rapid and accurate assessment of RV function in children with normal heart anatomy.</p> Graphical Abstract <p></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Feasibility and accuracy of the fully automated three-dimensional echocardiography right ventricular quantification software in children: validation against cardiac magnetic resonance

  • Qian Liu,
  • Zhiwei Zheng,
  • Yiwei Zhang,
  • Anjun Wu,
  • Jie Lou,
  • Xin Chen,
  • Yating Yuan,
  • Mingxing Xie,
  • Li Zhang,
  • Peng Sun,
  • Wei Sun,
  • Qing Lv

摘要

Background

Previous studies have confirmed that fully automated three-dimensional echocardiography (3DE) right ventricular (RV) quantification software can accurately assess adult RV function. However, data on its accuracy in children are scarce.

Objective

This study aimed to test the accuracy of the software in children using cardiac magnetic resonance (MR) as the gold standard.

Materials and methods

This study prospectively enrolled 82 children who underwent both echocardiography and cardiac MR within 24 h. The RV end-diastolic volume (EDV), end-systolic volume (ESV), and ejection fraction (EF) were obtained using the novel 3DE-RV quantification software and compared with cardiac MR values across different groups.

Results

The novel 3DE-RV quantification software was feasible in all 82 children (100%). Fully automated analysis was achieved in 35% patients with an analysis time of 8 ± 2 s and 100% reproducibility. Manual editing was necessary in the remaining 65% patients. The 3DE-derived RV volumes and EF correlated well with cardiac MR measurements (RVEDV, r=0.93; RVESV, r=0.90; RVEF, r=0.82; all <0.001). Although the automated approach slightly underestimated RV volumes and overestimated RVEF compared with cardiac MR in the entire cohort, the bias was smaller in children with RVEF ≥ 45%, normal RV size, and good 3DE image quality.

Conclusion

Fully automated 3DE-RV quantification software provided accurate and completely reproducible results in 35% children without any adjustment. The RV volumes and EF measured using the automated 3DE method correlated well with those from cardiac MR, especially in children with RVEF ≥ 45%, normal RV size, and good 3DE image quality. Therefore, the novel automated 3DE method may achieve rapid and accurate assessment of RV function in children with normal heart anatomy.

Graphical Abstract