Background <p>Photon-counting detector computed tomography (PCD-CT) can reduce radiation dose in paediatric lung imaging.</p> Objective <p>The aim of this study was to determine the lowest radiation dose maintaining adequate image quality for high-pitch lung imaging using a PCD-CT in a chest phantom replicating the characteristics of a 5-year-old child.</p> Materials and methods <p>The phantom was imaged on a dual-source PCD-CT with five different volume CT dose indices (CTDI<sub>vol</sub>): 0.45 mGy, 0.30 mGy, 0.15 mGy, 0.07 mGy, and 0.01 mGy. Scans were acquired with Sn100 kV in standard and ultra-high resolution modes. Polychromatic images were reconstructed with a 1-mm slice thickness, lung kernel Bl60, without quantum iterative reconstruction and with quantum iterative reconstruction at strengths 2 and 4. Two paediatric radiologists rated reconstructions subjectively, defining adequate image quality as the visibility of small peripheral structures. Objective evaluation included global noise index and global signal-to-noise ratio index.</p> Results <p>Exposure times were 0.42 s and 0.84 s for standard and ultra-high resolution modes, respectively. Subjective assessments showed no significant differences across scan modes or quantum iterative reconstruction strengths for both readers at all doses (all, <i>P</i> &gt; 0.05). Scans at 0.07 mGy with quantum iterative reconstruction 4 were deemed to maintain adequate image quality at the lowest dose. Global noise index was always lower and global signal-to-noise ratio index always higher in ultra-high resolution compared with standard mode, underscoring noise reduction achieved via ultra-high resolution mode’s small pixel effect.</p> Conclusions <p>PCD-CT enables high-pitch lung imaging while maintaining adequate image quality at a radiation dose as low as 0.07 mGy, with quantum iterative reconstruction 4, in a paediatric phantom representing a 5-year-old child.</p> Graphical Abstract <p></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Paediatric high-pitch lung imaging with photon-counting detector computed tomography: a dose reduction phantom study

  • Michael Zellner,
  • Thomas Sartoretti,
  • Thomas Flohr,
  • Thomas Frauenfelder,
  • Hatem Alkadhi,
  • Christian J. Kellenberger,
  • Victor Mergen

摘要

Background

Photon-counting detector computed tomography (PCD-CT) can reduce radiation dose in paediatric lung imaging.

Objective

The aim of this study was to determine the lowest radiation dose maintaining adequate image quality for high-pitch lung imaging using a PCD-CT in a chest phantom replicating the characteristics of a 5-year-old child.

Materials and methods

The phantom was imaged on a dual-source PCD-CT with five different volume CT dose indices (CTDIvol): 0.45 mGy, 0.30 mGy, 0.15 mGy, 0.07 mGy, and 0.01 mGy. Scans were acquired with Sn100 kV in standard and ultra-high resolution modes. Polychromatic images were reconstructed with a 1-mm slice thickness, lung kernel Bl60, without quantum iterative reconstruction and with quantum iterative reconstruction at strengths 2 and 4. Two paediatric radiologists rated reconstructions subjectively, defining adequate image quality as the visibility of small peripheral structures. Objective evaluation included global noise index and global signal-to-noise ratio index.

Results

Exposure times were 0.42 s and 0.84 s for standard and ultra-high resolution modes, respectively. Subjective assessments showed no significant differences across scan modes or quantum iterative reconstruction strengths for both readers at all doses (all, P > 0.05). Scans at 0.07 mGy with quantum iterative reconstruction 4 were deemed to maintain adequate image quality at the lowest dose. Global noise index was always lower and global signal-to-noise ratio index always higher in ultra-high resolution compared with standard mode, underscoring noise reduction achieved via ultra-high resolution mode’s small pixel effect.

Conclusions

PCD-CT enables high-pitch lung imaging while maintaining adequate image quality at a radiation dose as low as 0.07 mGy, with quantum iterative reconstruction 4, in a paediatric phantom representing a 5-year-old child.

Graphical Abstract