Background <p>Motion correction methods based on slice-to-volume registration (SVR) for fetal magnetic resonance imaging (MRI) allow reconstruction of three-dimensional (3-D) isotropic images of the fetal brain and body. However, all existing SVR methods are confined to research settings, which limits clinical integration. Furthermore, there have been no reported SVR solutions for low-field 0.55-T MRI.</p> Objective <p>Integration of automated SVR motion correction methods directly into fetal MRI scanning process via the Gadgetron framework to enable automated T2-weighted (T2W) 3-D fetal brain and body reconstruction in the low-field 0.55-T MRI scanner within the duration of the scan.</p> Materials and methods <p>A deep learning fully automated pipeline was developed for T2W 3-D rigid and deformable (D/SVR) reconstruction of the fetal brain and body of 0.55-T T2W datasets. Next, it was integrated into 0.55-T low-field MRI scanner environment via a Gadgetron workflow that enables launching of the reconstruction process directly during scanning in real-time.</p> Results <p>During prospective testing on 12 cases (22–40&#xa0;weeks gestational age), the fetal brain and body reconstructions were available on average 6:42 ± 3:13&#xa0;min after the acquisition of the final stack and could be assessed and archived on the scanner console during the ongoing fetal MRI scan. The output image data quality was rated as good to acceptable for interpretation. The retrospective testing of the pipeline on 83 0.55-T datasets demonstrated stable reconstruction quality for low-field MRI.</p> Conclusion <p>The proposed pipeline allows scanner-based prospective T2W 3-D motion correction for low-field 0.55-T fetal MRI via direct online integration into the scanner environment.</p> Graphical Abstract <p></p>

错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Scanner-based real-time three-dimensional brain + body slice-to-volume reconstruction for T2-weighted 0.55-T low-field fetal magnetic resonance imaging

  • Alena Uus,
  • Sara Neves Silva,
  • Jordina Aviles Verdera,
  • Kelly Payette,
  • Megan Hall,
  • Kathleen Colford,
  • Aysha Luis,
  • Helena Sousa,
  • Zihan Ning,
  • Thomas Roberts,
  • Sarah McElroy,
  • Maria Deprez,
  • Joseph Hajnal,
  • Mary Rutherford,
  • Lisa Story,
  • Jana Hutter

摘要

Background

Motion correction methods based on slice-to-volume registration (SVR) for fetal magnetic resonance imaging (MRI) allow reconstruction of three-dimensional (3-D) isotropic images of the fetal brain and body. However, all existing SVR methods are confined to research settings, which limits clinical integration. Furthermore, there have been no reported SVR solutions for low-field 0.55-T MRI.

Objective

Integration of automated SVR motion correction methods directly into fetal MRI scanning process via the Gadgetron framework to enable automated T2-weighted (T2W) 3-D fetal brain and body reconstruction in the low-field 0.55-T MRI scanner within the duration of the scan.

Materials and methods

A deep learning fully automated pipeline was developed for T2W 3-D rigid and deformable (D/SVR) reconstruction of the fetal brain and body of 0.55-T T2W datasets. Next, it was integrated into 0.55-T low-field MRI scanner environment via a Gadgetron workflow that enables launching of the reconstruction process directly during scanning in real-time.

Results

During prospective testing on 12 cases (22–40 weeks gestational age), the fetal brain and body reconstructions were available on average 6:42 ± 3:13 min after the acquisition of the final stack and could be assessed and archived on the scanner console during the ongoing fetal MRI scan. The output image data quality was rated as good to acceptable for interpretation. The retrospective testing of the pipeline on 83 0.55-T datasets demonstrated stable reconstruction quality for low-field MRI.

Conclusion

The proposed pipeline allows scanner-based prospective T2W 3-D motion correction for low-field 0.55-T fetal MRI via direct online integration into the scanner environment.

Graphical Abstract