Validation of a smartphone-compatible MIP-based sensor for bisphenol A determination in wastewater samples

  • Daria Yarynka,
  • Anton Honcharenko,
  • Larysa Gorbach,
  • Elena Piletska,
  • Sergey Piletsky,
  • Oleksandr Brovko,
  • Tetyana Sergeyeva

摘要

A handheld smartphone-compatible molecularly imprinted polymer (MIP)-based sensor was developed for the analysis of bisphenol A (BPA) in wastewater samples. Sensing elements based on ethylene glycol methacrylate phosphate (EGMP)-containing MIP films were designed and optimized using molecular dynamics simulations. The highly porous MIP films were synthesized via in situ polymerization, employing a fragment-based approach. The colorimetric response was based on the 4-aminoantipyrine method, while the MIP films were further utilized to detect BPA with a smartphone. The proposed sensor exhibited a wide linear range from 5 to 250 μM, with a limit of detection (LOD) of 5 μM (S/N = 3). Furthermore, the designed analytical system demonstrated excellent analytical performance in terms of selectivity, stability, and reproducibility. During sensor validation, real wastewater samples were successfully tested for BPA, showcasing the feasibility of the smartphone-compatible MIP-based sensor. Recovery values of 87.1–114.6% underscored the efficacy and reliability of the developed sensor system.