Measurement of Healthy Adult Brain Temperature Using 1H Magnetic Resonance Spectroscopy Thermometry
摘要
The purpose of this study is to measure the brain temperature (Tbr) by using 1H magnetic resonance spectroscopy (1H MRS) thermometry and investigate its age and gender differences in healthy adults. The brain temperature was further compared with the body temperature (Tbo) to investigate the possible existence of brain-body temperature gradient (∆T).
MethodsA total of 80 subjects were included in this study. 1H MRS data were collected on a 3.0T MR scanner using Point Resolved Selective Spectroscopy (PRESS) sequence. Voxels were positioned in the right frontal (RF) lobe and left frontal (LF) lobe, respectively. The temperature of each voxel was calculated by chemical shift difference (∆δ) between H2O and NAA which was obtained by LCModel software. The average temperature of bilateral frontal lobe voxels was defined as Tbr for each subject. The average forehead temperature was acquired before MR scanning, defined as Tbo, in this study. The difference between Tbr and Tbo, denoted as the brain-body temperature gradient (∆T), was calculated. Age and gender characteristics of Tbr, ∆T and Tbo were analyzed.
ResultsTbr (38.51 ± 0.59℃) was higher than Tbo (36.47 ± 0.26℃) (P < 0.05). Negative correlations were observed between Tbr and age (r = −0.49, P < 0.05) and between ∆T and age (r = −0.44, P < 0.05), whereas no correlation existed between Tbo and age (r = −0.03, P = 0.79).
ConclusionOur observation demonstrated that the brain temperature, derived from 1H MRS thermometry, is significantly higher than the body temperature, indicating the existence of a brain-body temperature gradient, and the brain temperature gradually decreases with age.