错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

The DERIVO 2 Heal Embolization Device in the Treatment of Ruptured and Unruptured Intracranial Aneurysms: a Retrospective Multicenter Study

  • Roland Schwab,
  • Christoph Kabbasch,
  • Lukas Goertz,
  • Marius Kaschner,
  • Daniel Weiss,
  • Christian Loehr,
  • Hauke Wensing,
  • Maxim Bester,
  • Andreas Simgen,
  • André Kemmling,
  • Christina Wendl,
  • Erelle Fuchs,
  • Maximilian Thormann,
  • Daniel Behme,
  • Hannes Nordmeyer

摘要

Backround

The use of flow diverting stents in the treatment of intracranial aneurysms is associated with a risk of neurological morbidity due to their thrombogenicity. To reduce this risk different surface modifications have been developed. The Derivo 2 Embolization Device (Acandis, Pforzheim, Germany) has proven to be a safe and effective flow diverter. To overcome the risk of thrombo-embolism, the device was modified by adding an anti-thrombogenic fibrin-heparin coating. We aimed to assess the safety and effectiveness of the Derivo 2 heal Embolization Device.

Methods

Retrospective multicenter data from nine German neurovascular centers between February 2022 until December 2023 were used. Patients treated with the Derivo 2 heal Embolization Device for unruptured or ruptured intracranial aneurysms were included. Peri- and postprocedural adverse events, clinical outcomes, and angiographic follow-up results were evaluated.

Results

84 patients (73.8% female; mean age 58.7 years) with 89 aneurysms (mean size 9.8 mm) were included. 87.6% were located in the anterior circulation. Most of them were sidewall aneurysms (88.8%). 96 flow diverters were used. 99.0% were successfully implanted. An in-stent balloon angioplasty was performed in 6.0% of the cases. An additional coiling was performed in 28.6%. Technical difficulties were present in 12.0% of the cases. Thrombotic events occurred in 4.8% with no neurological sequelae. Mortality and morbidity were 0 and 1.2% respectively. Adequate aneurysm occlusion was achieved in 80.7% with a mean follow-up time of 6.6 months.

Conclusion

The Derivo 2 heal Embolization Device showed a satisfying aneurysm occlusion and safety with a low rate of neurological morbidity.