Long-term Safety and Efficacy of the Derivo Embolization Device in a Single-center Series
摘要
This study analyzes the long-term clinical and angiographic outcomes of the Derivo Embolization Device (DED), an advanced flow diverter device with an electropolished surface, for the treatment of intracranial aneurysms.
MethodsA consecutive series of 101 patients (mean age: 58 years, 72% female) treated with the DED for 122 aneurysms at a single center between 2017 and 2023 was retrospectively analyzed for major (change in National Institutes of Health Stroke Scale [NIHSS] score ≥ 4 points) and minor (change in NIHSS score < 4 points) neurological events, procedural morbidity (increase of at least one point on the modified Rankin Scale), and angiographic results.
ResultsThere were 14 (11%) recurrent aneurysms, 15 (12%) ruptured aneurysms, 26 (21%) posterior circulation aneurysms and 16 (13%) fusiform or dissecting aneurysms. Device deployment failed in 1 case (1%). Procedure-related symptomatic procedural complications consisted of 2 (2%) major events (1 major stroke and 1 vessel perforation with intracranial hemorrhage and infarction) and 6 minor events (6 minor strokes). Procedural morbidity was 5%. There were no late ischemic or hemorrhagic events during follow-up. Complete and favorable aneurysm occlusion was achieved in 54% (40/74) and 62% (46/74) at a mean of 5 months, 71% (27/38) and 87% (33/38) at a mean of 12 months, and 76% (25/33) and 97% (32/33) at a mean of 35 months, respectively.
ConclusionThe results demonstrate progressive aneurysm occlusion beyond 12 months after DED implantation with an almost 100% favorable occlusion rate. Procedural morbidity was low and there were no late complications.