错误:搜索内容不能为空,请输入英文关键词
错误:关键词超出字数限制,请精简
高级检索

Safety and Efficacy of LVIS Jr Stent-assisted Coiling of Intracranial Aneurysms in Small-diameter Parent Arteries

  • Shuailong Shi,
  • Shuhai Long,
  • Fangfang Hui,
  • Qi Tian,
  • Zhuangzhuang Wei,
  • Ji Ma,
  • Jie Yang,
  • Ye Wang,
  • Xinwei Han,
  • Tengfei Li

摘要

Objective

To investigate the safety and efficacy of LVIS Jr stent-assisted coiling (SAC) of intracranial aneurysms (IAs) in small-diameter parent arteries and determine the factors influencing incomplete aneurysm occlusion.

Material and Methods

Clinical and imaging data of 130 patients with IAs in small-diameter parent arteries that were treated with LVIS Jr SAC were retrospectively analyzed. Stent apposition was evaluated by high-resolution flat detector CT, and aneurysm embolization density was evaluated using 2D-DSA. Perioperative complications were recorded. Multivariate logistic regression analyses were performed to determine possible factors for incomplete aneurysm occlusion.

Results

In this study, 130 patients (60 and 70 patients with ruptured and unruptured aneurysms, respectively) were successfully treated with LVIS Jr SAC. Immediate digital subtraction angiography (DSA) showed that the aneurysm occlusion was Raymond-Roy class I, II, IIIa, and IIIb in 93 (71.5%), 24 (18.5%), 8 (6.2%), and 5 (3.8%) cases, respectively. There were three cases of acute in-stent thrombosis and two cases of severe vasospasm observed during the perioperative period. The 6‑month follow-up angiograms indicated that complete aneurysm occlusion in 122 patients was 79.5% (97/122). Multivariate logistic regression analyses showed that an aneurysm size > 10.0 mm, parent artery mean diameter < 2.0 mm, and incomplete stent apposition at the aneurysm neck were possible risk factors for incomplete aneurysm occlusion.

Conclusion

The LVIS Jr SAC is effective for managing IAs in small-diameter parent arteries. An aneurysm size > 10.0 mm, parent artery mean diameter < 2.0 mm, and incomplete stent apposition at the aneurysm neck are possible risk factors for incomplete aneurysm occlusion.