Is bracket bonding with guided bonding devices accurate enough for crowded dentition?

  • Peiqi Wang,
  • Bin Li,
  • Yuting Tang,
  • Yixi Huang,
  • Xianglong Han,
  • Ding Bai,
  • Chaoran Xue

摘要

Background

This research aimed to study whether bracket bonding using guided bonding devices (GBDs) is accurate enough for crowded dentitions in vitro.

Methods

Fifteen three-dimensionally (3D) printed resin model sets were included and divided into three groups: mild, moderate, and severe crowding. The resin models were scanned and virtually bonded with brackets. Corresponding GBDs were generated and 3D printed. Subsequently, the brackets were bonded to the resin models on a dental mannequin using the GBDs. The models with bonded brackets were scanned, and comparisons were made between the positions of the actually bonded and the planned ones to evaluate possible deviations.

Results

There was no immediate bonding failure in any group. The bonding duration tended to increase with crowding severity (P > 0.05). Almost all linear and angular deviations in all groups were below 0.5 mm and 2°, respectively, and no statistically significant difference was found among the different crowding degrees (P > 0.05). In all groups, the brackets tended to deviate lingually and had buccal crown torque. Brackets in the groups with mild and severe crowding showed a tendency for mesiobuccal rotation.

Conclusion

GBDs provide high bracket bonding accuracy for dentitions with different crowding degrees and, thus, could hopefully be applied to uncrowded and crowded dentitions alike.